• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SURGICAL, INC. ZIMMER AIR DERMATOME

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL, INC. ZIMMER AIR DERMATOME Back to Search Results
Catalog Number 00880100100
Device Problem Power Problem (3010)
Patient Problem No Patient Involvement (2645)
Event Date 11/05/2019
Event Type  malfunction  
Manufacturer Narrative
This event has been recorded under zimmer biomet complaint number cmp- the customer returned an air dermatome device, serial number (b)(4), for evaluation.The customer also returned a hose and 1''/2''/3''/4'' width plates, for evaluation.Product review of the air dermatome on november 6, 2019 revealed that motor speed was below specifications.The calibration was within specifications and the control bar was in the correct position.Repair of the air dermatome was performed by zimmer biomet surgical on november 6, 2019 which included replacement of the hose, poppet spring, o-rings, wave washers, o-ring, motor sleeve, motor, needle bearing, throttle lever, reciprocating arm, spring seal, semi-circle bearings, and vespel bearings.Air dermatome, serial number (b)(4), was then tested and functioned properly.It was repaired, inspected and tested.While the returned product investigation confirmed that the air dermatome had a malfunctioning motor, it cannot be determined from the information provided what actually caused the reported event.Therefore, based on the information provided, a specific root cause of the reported event cannot be determined.The device was noted to be functioning as intended after the hose, poppet spring, o-rings, wave washers, o-ring, motor sleeve, motor, needle bearing, throttle lever, reciprocating arm, spring seal, semi-circle bearings, and vespel bearings were replaced.The investigation was based on the information that was provided initially and any information that was obtained throughout the follow-up process.
 
Event Description
It was reported that the device required pm with no alleged deficiency.The investigation identified that the motor speed was below specifications.No adverse events were reported as a result of this malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZIMMER AIR DERMATOME
Type of Device
DERMATOME
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
christina arnt
56 e. bell drive
warsaw, IN 46582
5745273773
MDR Report Key9418959
MDR Text Key197752813
Report Number0001526350-2019-01120
Device Sequence Number1
Product Code GFD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial
Report Date 12/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/05/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number00880100100
Device Lot Number62682956
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/05/2019
Was the Report Sent to FDA? No
Date Manufacturer Received11/21/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/25/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
-
-