• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ATTUNE DISTAL FEM CUT BLOCK; ATTUNE INSTRUMENTS : RESECTION BLOCKS/DEVICES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ATTUNE DISTAL FEM CUT BLOCK; ATTUNE INSTRUMENTS : RESECTION BLOCKS/DEVICES Back to Search Results
Catalog Number 254400522
Device Problems Connection Problem (2900); Material Deformation (2976); Physical Resistance/Sticking (4012)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/12/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that the pin binds up in the hole.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary : examination of the returned device confirmed the reported event.The noted damage is consistent with device wear out due to normal intended use and the investigation did not establish a need for corrective action.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE DISTAL FEM CUT BLOCK
Type of Device
ATTUNE INSTRUMENTS : RESECTION BLOCKS/DEVICES
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
MDR Report Key9419092
MDR Text Key169730603
Report Number1818910-2019-121048
Device Sequence Number1
Product Code HTZ
UDI-Device Identifier10603295423737
UDI-Public10603295423737
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 11/12/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number254400522
Device Lot NumberABC84145
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/10/2019
Initial Date Manufacturer Received 11/12/2019
Initial Date FDA Received12/05/2019
Supplement Dates Manufacturer Received12/10/2019
01/02/2020
Supplement Dates FDA Received12/18/2019
01/06/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-