Model Number 3-SPIKE DISPOSABLE SET
Device Problem Temperature Problem (3022)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/23/2019
Event Type  malfunction  
Manufacturer Narrative
The implicated disposable set has not been returned for investigation, nor was the serial number of the rapid infuser reported.The manufacturing records for this lot were reviewed and no anomalies were identified.When the rapid infuser recognizes a situation that may compromise safe and effective infusion, the system stops pumping and heating, closes off the line to the patient, sounds an audible alarm, and displays an alarm message with instructions for corrective measure.No patient injury was reported.The investigation is not yet complete.A follow-up report will be submitted.
 
Event Description
Belmont's distributor received a complaint from the user facility and relayed the following report: "over temperature".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THE BELMONT RAPID INFUSER
Type of Device
THERMAL INFUSION FLUID WARMER
Manufacturer (Section D)
BELMONT MEDICAL TECHNOLOGIES
780 boston road
billerica MA 01821
Manufacturer (Section G)
BELMONT MEDICAL TECHNOLOGIES
780 boston road
billerica MA 01821
Manufacturer Contact
sabrina belladue
780 boston road
billerica, MA 01821
9783307637
MDR Report Key9434257
MDR Text Key194995348
Report Number1219702-2019-00097
Device Sequence Number1
Product Code LGZ
UDI-Device Identifier00896128002022
UDI-Public00896128002022
Combination Product (y/n)N
Reporter Country CodeKS
PMA/PMN Number
K141654
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,user facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 12/06/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2022
Device Model Number3-SPIKE DISPOSABLE SET
Device Catalogue Number903-00006P
Device Lot Number2019-04 09
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 11/06/2019
Initial Date FDA Received12/06/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/01/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-