Model Number N/A
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Host-Tissue Reaction (1297); Ossification (1428); Bone Fracture(s) (1870); Pain (1994); Osteopenia/ Osteoporosis (2651); No Code Available (3191)
Event Date 04/26/2019
Event Type  Injury  
Manufacturer Narrative
(b)(4).Concomitant medical products: item# 11-363662 36mm cocr mod hd std lot# 618180, item# ep-108323 e-poly 36mm +3 hiwall lnr sz23 lot# 619330, item# pt-116052 regen/rnglc+ ltd 52mm sz 23 lot# 480870.(b)(6).Customer has indicated that the product will not be returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001825034 - 2019 - 05488.
 
Event Description
It was reported a patient underwent right hip arthroplasty.Subsequently, the patient was revised approximately 5 years post initial implantation due to pain, heterotopic ossification, pseudocapsule, periprosthetic fracture, poor bone quality, and implant (cup) malposition.Attempts were made to obtain additional information; however, none was available.
 
Manufacturer Narrative
(b)(4).Reported event was confirmed by review of medical records.Device history record (dhr) was reviewed and no discrepancies were found.Root cause was unable to be determined.Medical records indicate right hip pain.She had signs of stress reaction around the stem.Present was heterotopic ossification and malposition of her acetabular component.No obvious signs of infection.Several cultures were sent to microbiology (no report).A pseudocapsulotomy was performed.Femoral stem was well fixed.It was removed without iatrogenic bone loss.The patient¿s lateral cortex was found to be thin.Upon maneuvering the hip, a non-displaced fracture of the greater trochanter just below the vastus ridge was identified.Extensive heterotopic ossification was noted and removed.Acetabulum was found to be vertical and retroverted.The acetabulum was removed without iatrogenic bone loss.There was a silver dollar sized medial wall defect.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TAPERLOC MICROP LAT FMRL 9.0MM
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key9435207
MDR Text Key169796006
Report Number0001825034-2019-05487
Device Sequence Number1
Product Code PBI
Combination Product (y/n)N
PMA/PMN Number
K121874
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,health professional
Type of Report Initial,Followup
Report Date 03/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number15-103203
Device Lot Number050760
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/15/2019
Initial Date FDA Received12/08/2019
Supplement Dates Manufacturer Received03/06/2020
Supplement Dates FDA Received03/13/2020
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-