Catalog Number 3003940001 |
Device Problem
Unsealed Device Packaging (1444)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 11/27/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Report source, foreign - event occurred in (b)(6).The device was not returned to the manufacturer.Therefore it could not be analyzed.The review of the device manufacturing quality record indicate that (b)(4) products désignation refobacin bone cement r 40x1, reference 3003940001, batch a744da0211 were manufactured on (12th april 2018).The device manufacturing quality record (b)(4) indicates that the released product met all requirements to perform as intended.No non conformity or deviation was observed which could be linked to the complaint issue (complaint category inner cement pouch open sealing) event described in the complaint.The device manufacturing quality record indicates that the released product met all requirements to perform as intended.According to available data, the most probable root cause is due to packaging issue (sealing process).Corrective action has been initiated to address reported issue.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
It was reported that the inner sterile packaging is opened during the operation, use spare bone cement complete the operation.No adverse events have been reported as a result of the malfunction.
|
|
Manufacturer Narrative
|
(b)(4).This follow-up report is being submitted to relay additional information.The attached photo confirms the reported event; the inner cement pouch is opened.The product analysis can't be performed as the product was not returned.The review of the device manufacturing quality record indicates that (b)(4) designation refobacin bone cement r 40x1, reference 3003940001, batch a744da0211 were manufactured on (12th april 2018).The device manufacturing quality record (of6279719) indicates that the released product met all requirements to perform as intended.No non conformity or deviation was observed which could be linked to the complaint issue (complaint category inner cement pouch open sealing) event described in the complaint.1 complaint has been recorded over the batch number a744da0211.According to the available data, the most probable root cause is due to packaging issue (sealing process).Corrective action has been initiated to address reported issue.A customer letter will be sent to the complainant in order to explain the issue.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.
|
|
Event Description
|
It was reported that the inner sterile packaging is opened during the operation, use spare bone cement complete the operation.No adverse events have been reported as a result of the malfunction.
|
|