Reportable based on device analysis completed on 03dec2019.It was reported that the balloon could not inflate.A 10/2.50 flextome cutting balloon was selected for use.During the procedure, the balloon was able to reach the lesion; however, it was noted that the balloon could not inflate.The procedure was completed with another of same the device.No patient complications were reported and the patient was stable post procedure.However, device analysis revealed that there was a balloon pinhole detected 3mm distal of the distal edge of the proximal markerband.
|