MAUDE Adverse Event Report: ABBOTT IRELAND DIAGNOSTICS DIVISION LONGFORD ARCHITECT FREE T3; FREE TRIIODOTHYRONINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT IRELAND DIAGNOSTICS DIVISION LONGFORD ARCHITECT FREE T3; FREE TRIIODOTHYRONINE Back to Search Results
Catalog Number 07K63-37
Device Problem Low Test Results (2458)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Patient information, patient identifier: multiple = sid (b)(6).Product correction letters were issued on 12oct2018 to all architect total t3 (tt3) and free t3 (ft3) customers, using the i1000sr, i2000, or i2000sr instruments and who received one of the impacted lots.The letter instructed the customers to prevent this interaction by either separating the architect ft3 and/or architect tt3 assays by running these tests on different instruments from the following assays [architect tsh, architect t-uptake, architect hiv ag/ab combo, architect cortisol, architect lh, architect progrp, architect rhtlv-i/ii, architect total psa, architect afp, and architect free psa, architect 25-oh vitamin d], or by performing daily maintenance on the instrument prior to performing batch testing for all ft3 and/or tt3 samples.The mechanism of carryover involves carryover of poly-l-lysine (pll) from other assays into the conjugate dispense step in the ft3 and tt3 assays through adherence of the pipetting probe.Correction/removal number = 3005094123-10/22/18-001-c.
 
Event Description
The customer observed falsely depressed architect free t3 results for 9 patients, while comparing results between 2-point calibrator and 6-point calibrator.The customer noticed that the generated results were inconsistent with tsh/ft4 results.The following data was generated: sid (b)(6): 2.56 (current result), 2.88 (previous result), 2.064 (tsh), and 12.27 (free t4).Sid (b)(6): <2.3 (current result), 2.86 (previous result), 45.216 (tsh), and 5.15 (free t4).Sid 51319763: 2.47 (current result), 3.82 (previous result), 1.67 (tsh), and 17.07 (free t4).Sid (b)(6): 2.43 (current result), 3.58 (previous result), 0.39 (tsh), and 12.64 (free t4).Sid (b)(6): 2.47 (current result), 4.25 (previous result), 1.329 (tsh), and 9.63 (free t4).Sid (b)(6): 2.37 (current result), 3.36 (previous result), 0.385 (tsh), and 11.71 (free t4).Sid (b)(6): 2.55 (current result), 3.23 (previous result), 0.13 (tsh), and 15.67 (free t4).Sid (b)(6): <2.33 (current result), 2.88 (previous result), 0.425 (tsh), and 11.91 (free t4).Sid (b)(6): 2.56 (current result), 4.31 (previous result), 0.232 (tsh), and 12.46 (free t4).There was no reported impact to patient management.
 
Manufacturer Narrative
Upon further review, the customer's issue is not associated with remedial action (3005094123-10/22/18-001-c) and is unassigned from the field action.Through the review of this complaint, it was identified that the issue is not related to fa12oct2018a, therefore, this complaint is no longer a reportable event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCHITECT FREE T3
Type of Device
FREE TRIIODOTHYRONINE
Manufacturer (Section D)
ABBOTT IRELAND DIAGNOSTICS DIVISION LONGFORD
lisnamuck
co. longford
longford NA
EI  NA
MDR Report Key9437758
MDR Text Key219780058
Report Number3005094123-2019-00356
Device Sequence Number1
Product Code CDP
UDI-Device Identifier00380740144838
UDI-Public00380740144838
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Remedial Action Notification
Type of Report Initial,Followup
Report Date 02/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/28/2020
Device Catalogue Number07K63-37
Device Lot Number02124UI00
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 11/25/2019
Initial Date FDA Received12/09/2019
Supplement Dates Manufacturer Received02/19/2020
Supplement Dates FDA Received02/28/2020
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction Number3005094123-10/22/18-001-
Patient Sequence Number1
Treatment
ARCHITECT I2000SR ANALYZER; ARCHITECT I2000SR ANALYZER, LN 03M74-02,; LIST 03M74-02, SERIAL (B)(6); SERIAL # (B)(6)
-
-