BOSTON SCIENTIFIC CORPORATION WOLVERINE CORONARY CUTTING BALLOON MONORAIL; CATHETER, PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY (PTCA), CUTTING/SCORING Back to Search Results
Model Number 3851
Device Problem Inflation Problem (1310)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/29/2019
Event Type  malfunction  
Manufacturer Narrative
Device evaluated by manufacturer: the device was returned for analysis.A visual examination identified that the balloon was not folded which indicates that the balloon was subjected to positive pressure.Blood was identified within the balloon and lumen which is evidence of a device leak.The returned device was attached to a boston scientific encore inflation unit and positive pressure was applied using glycerol/water in an attempt to inflate the balloon.The balloon could not be inflated due to the presence of solidified media that was present within the inflation lumen.The device was soaked in a water bath to help soften the media before further inflation attempts were made.The device was removed from the bath and the balloon was again attached to an inflation device subjected to positive pressure and liquid was observed to be leaking from a balloon pinhole located approximately 1mm proximal of the proximal markerband.An examination of the balloon material and proximal markerband identified no issues which could potentially have contributed to this complaint.The markerbands, blades and tip section of the device were visually and microscopically examined, and no issues were noted that may have potentially contributed to the complaint incident.All blades were present and fully bonded to the balloon material.A visual and tactile examination found the hypotube of the device to be kinked at more than one location.This type of damage is consistent with excessive force being applied to the device.No other issues were identified during the product analysis.
 
Event Description
Reportable based on device analysis completed on 28nov2019.It was reported that balloon could not be inflated.A 10mmx3.00mm wolverine coronary cutting balloon monorail was selected for use.During procedure, it was noted that the balloon could not be inflated.The procedure was completed with another of the same device.No patient complications were reported.However, device analysis revealed a balloon pinhole.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WOLVERINE CORONARY CUTTING BALLOON MONORAIL
Type of Device
CATHETER, PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY (PTCA), CUTTING/SCORING
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC IRELAND LIMITED
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key9439378
MDR Text Key176020703
Report Number2134265-2019-15243
Device Sequence Number1
Product Code NWX
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P950020
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Physician
Type of Report Initial
Report Date 12/09/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/09/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3851
Device Catalogue Number3851
Device Lot Number0024136046
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/11/2019
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/28/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/19/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-