Model Number 319.006
Device Problem Break (1069)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
Product complaint #: (b)(4).Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Without a lot number, the device history records review could not be completed.The investigation could not be completed; no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that on (b)(6) 2019, three (3) 2.0mm/2.4mm depth gauge were broken, one (1) 1.3mm/1.5mm depth gauge was damaged, and one (1) reduction forcep with point speed lock was damaged.The issue was discovered at sterile processing department.There was no patient involvement.This complaint involves three (3) devices.This report is 3 of 3 for (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.E1: hospital phone number added.E1: reporter is a synthes employee.H6: part number: 319.006; synthes lot number: ft00309; supplier lot number: n/a; release to warehouse date: 6jan2017; expiration date: n/a; supplier: (b)(4).No ncrs were generated during production.Review of the device history record showed that there were no issues during the manufacture of this product, and any subcomponents, which would contribute to this complaint condition.Visual inspection: it was notice the device's needle was completely broken off.Hence, the complaint is confirmed.The break occurred where the needle threads into the slider: the needle broke flush with the slider.The broken needle was not returned.Dimensional inspection: dimensional inspection of needle component could not be conducted due to the post manufacturing deformation at the break, the location of the break, and since the distal portion of the needle was not received.Investigation conclusion: the exact cause of the complaint condition cannot be determined as the handling and use of the device are unknown.However, the condition of the depth gauge is consistent with the result of a bending force applied to the needle portion of the depth gauge that is beyond the yield limit of the material.During the investigation, no product design or manufacturing issues were observed that may have contributed to the complaint condition; therefore, further corrective and/or preventive action is not required.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEPTH GAUGE FOR 2.0MM AND 2.4MM SCREWS
Type of Device
GAUGE,DEPTH
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key9440316
MDR Text Key196419580
Report Number2939274-2019-62602
Device Sequence Number1
Product Code HTJ
UDI-Device Identifier10886982189943
UDI-Public(01)10886982189943
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 11/19/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number319.006
Device Catalogue Number319.006
Device Lot NumberFT00309
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/12/2019
Initial Date Manufacturer Received 11/19/2019
Initial Date FDA Received12/09/2019
Supplement Dates Manufacturer Received12/30/2019
Supplement Dates FDA Received01/08/2020
Patient Sequence Number1
-
-