DEPUY SYNTHES PRODUCTS LLC BATTERY OSCILLATOR II FOR BPL II; MOTOR, SURGICAL INSTRUMENT, AC-POWERED Back to Search Results
Catalog Number 530.710
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/01/2019
Event Type  malfunction  
Manufacturer Narrative
The actual device has been returned and is currently pending evaluation.Once the evaluation of the device has been completed, a supplemental medwatch report will be sent accordingly.Udi: (b)(4).
 
Event Description
It was reported from (b)(6) that the screw of the battery oscillator device, under the blade fixing part, was removed.It was not reported if the event occurred during a surgical procedure.It was not reported if there was a delay to a planned procedure.It was not reported if there was a spare device available for use.There was no patient involvement reported.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was not reported, however, it was reported that the event occurred in 2019.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Manufacturer Narrative
H10: depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.The actual device was returned for evaluation.The battery oscillator device was evaluated and the reported condition that the screw under the blade fixing part was removed was not confirmed.Therefore, an assignable root cause was not determined.However, during evaluation it was observed that the saw head locking mechanism was cracked.It was determined that the root cause for the cracked oscillation saw head was covered under a capa.The assignable root cause was determined to be traced to component failure.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BATTERY OSCILLATOR II FOR BPL II
Type of Device
MOTOR, SURGICAL INSTRUMENT, AC-POWERED
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
eimattstrasse 3
oberdorf 4436
SZ  4436
MDR Report Key9440425
MDR Text Key195537346
Report Number8030965-2019-70847
Device Sequence Number1
Product Code GEY
UDI-Device Identifier07611819491847
UDI-Public7611819491847
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 11/21/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number530.710
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/22/2019
Initial Date Manufacturer Received 11/21/2019
Initial Date FDA Received12/09/2019
Supplement Dates Manufacturer Received01/16/2020
Supplement Dates FDA Received01/23/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-