Model Number CI-1601-04
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problems Unspecified Infection (1930); Swelling (2091)
Event Date 11/11/2019
Event Type  Injury  
Event Description
The recipient reportedly experienced an ear infection and was placed on antibiotics, but the infection spread causing the ear to swell.The recipient's device was explanted.The recipient will be reimplanted at a later date.The recipient was placed on iv antibiotics.
 
Manufacturer Narrative
Advanced bionics considers the investigation into this reportable event as closed.The recipient reportedly recovered.The external visual inspection revealed the electrode was severed near the electrode ground ring prior to receipt.This is believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock was verified.The electrode condition prevented an electrical test from being performed.The device passed the electrical and mechanical tests performed.This device was explanted for medical reasons.The device passed the tests performed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIRES¿ ULTRA 3D IMPLANT
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
ADVANCED BIONICS, LLC
28515 westinghouse place
valencia CA 91355
MDR Report Key9440887
MDR Text Key170010268
Report Number3006556115-2019-00725
Device Sequence Number1
Product Code MCM
UDI-Device Identifier07630016862250
UDI-Public(01)07630016862250(11)180904(17)210831
Combination Product (y/n)N
PMA/PMN Number
P960058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 11/12/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/09/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date08/31/2021
Device Model NumberCI-1601-04
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/25/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age29 MO
-
-