Reportable based on device analysis completed on 21nov2019.A 10mmx3.25mm wolverine coronary cutting balloon was selected for use.During the procedure, when the balloon was inflated, it was noted that the proximal side did not expand.Although pressure was applied up to the rated pressure, still inflation failure occurred.The procedure was completed with the original device.There were no complications reported.However, device analysis revealed that liquid was observed to be leaking from a longitudinal tear in the balloon.
|