ADVANCED BIONICS, LLC HIRES 90K IMPLANT; COCHLEAR IMPLANT Back to Search Results
Model Number CI-1400-02H
Device Problems Mechanical Problem (1384); Defective Device (2588)
Patient Problem Failure of Implant (1924)
Event Date 11/11/2019
Event Type  malfunction  
Event Description
The recipient reportedly experienced open electrodes.External equipment was exchanged and programming adjustments were made, however, the issue did not resolve.The recipient's device was explanted.The recipient was reimplanted with another advanced bionics cochlear device.
 
Manufacturer Narrative
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Manufacturer Narrative
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed that the array was severed prior to receipt.This is believed to have occurred during revision surgery.The photographic imaging inspection revealed broken electrode wires at the fantail region.System lock was verified.The electrode condition prevented some of the electrical tests from being performed.The device passed the electrical and mechanical tests performed.The scanning electron microscopy analysis of the electrode revealed evidence of fatigue breaks.The photographic inspection and scanning electron microscopy analysis revealed all broken electrode wires in the fantail region.It is believed that these breaks caused the multiple electrodes with open circuits.A corrective action was implemented.This is the final.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIRES 90K IMPLANT
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
ADVANCED BIONICS, LLC
28515 westinghouse place
valencia CA 91355
MDR Report Key9444774
MDR Text Key176491318
Report Number3006556115-2019-00730
Device Sequence Number1
Product Code MCM
Combination Product (y/n)N
PMA/PMN Number
P960058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup,Followup
Report Date 11/15/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date05/01/2011
Device Model NumberCI-1400-02H
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/14/2020
Initial Date Manufacturer Received Not provided
Initial Date FDA Received12/10/2019
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Supplement Dates FDA Received09/04/2020
02/02/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age12 YR
-
-