Catalog Number UNK HIP ACETABULAR LINER
Device Problem Naturally Worn (2988)
Patient Problems Osteolysis (2377); No Code Available (3191)
Event Date 11/21/2011
Event Type  Injury  
Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.(b)(4).
 
Event Description
The literature article entitled, ""femoral stem fracture and in vivo corrosion of retrieved modular femoral hips"" written by j.Caitlin huot carlson, ms, douglas w.Van citters, phd, john h.Currier, ms, amber m.Bryant, be, michael b.Mayor, md, and john p.Collier, de published by the journal of arthroplasty vol.27 no.7 2012 accepted by publisher 21 november 2011 was reviewed.The article's purpose was to report on retrieved modular femoral stems and inspect for corrosion.All stems were srom stems.Article reports that femoral heads were ceramic or metal.Liners may have been metal, ceramic or poly and the cups are not identified but assumed to be depuy implants.Inspection revealed that most stems had some degree of 'fretting' or corrosion at modular junctions including stem and sleeve interface or (and) stem and head junction.However the article does not quantify the findings to modular junction locations in order to determine accurate quantities.Table 1 provided further detail of reasons stems were explanted.Each case is captured individually for reason of revision and any information provided in the table and article in linked complaints.This complaint captures case no 26 that received revision when the stem was explanted at 113 months post initial implant for reason of osteolysis associated with liner (no information provided for debris or wear).The patient¿s age was 87 and gender identity female.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: product complaint # (b)(4).Investigation summary
=
> no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.Device history lot - null.Device history batch - null.Device history review - null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN HIP ACETABULAR LINERS
Type of Device
HIP ACETABULAR LINERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key9446362
MDR Text Key185226013
Report Number1818910-2019-122121
Device Sequence Number1
Product Code KWY
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,literatur
Type of Report Initial,Followup
Report Date 11/17/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNK HIP ACETABULAR LINER
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 11/17/2019
Initial Date FDA Received12/10/2019
Supplement Dates Manufacturer Received01/08/2020
Supplement Dates FDA Received01/13/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age27 YR
-
-