Catalog Number 311.34
Device Problems Corroded (1131); Contamination /Decontamination Problem (2895)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Device report from synthes reports an event in (b)(6) as follows: it was reported that during the inspection of the loan kit, it was discovered that the kit was wet and there was a rust in three devices.There was no patient involvement reported.This report is for one (1) tap for 4.0mm cancellous bone screws.This is report 1 of 1 for complaint (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Concomitant medical products: tap f/cancellousscr ø4 l110/85 (part # 311.340, lot # l643580, quantity 1).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h4, h6: based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.H3, h4, h6: reviewing attached picture, the complaint description can be confirmed that on the article some rusty spots are visible.Product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device history lot part: 311.340, lot: 9894189, manufacturing site: bettlach, release to warehouse date: 02.June 2016.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h4, h6: a review of the device history record.Device history lot part: 311.340, lot: 9894189, manufacturing site: bettlach, release to warehouse date: 02.June 2016.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.H3, h6: investigation summary reviewing attached picture, the complaint description can be confirmed that on the article some rusty spots are visible.Product was not returned.Based on the information available, it has been determined that no corrective and preventive action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H3, h4, h6 investigation summary.Visual inspection: the returned device was examined and minor wear and discoloration, drop-shaped spots, consistent with corrosion was noted on the device body.The observed damage is consistent with the reported complaint condition as such the complaint condition can be confirmed.All features related to the reported complaint condition were reviewed and no other issues were identified.Document/specification review: the manufacturing review shows that the production procedure was according to the specifications and there were no issues that would contribute to this complaint condition.Furthermore, as indicated in the manufacturing documents, the correct material (1.4112) was used according iso 7153-1.Summary: the received condition of the device is concordant with the reported complaint description and the complaint condition is therefore confirmed.However, since no fault could be found in material or during the production procedures.This failure is typically consistent with inadequate sterilization practices which allow residual moisture to remain on the device.Please note: regarding corrosion it can be stated, that all materials, including so called stainless steel as well, are conditionally only rust-resistant.The corrosion resistance is maintained only when the material is stored on a dry and metallic contactless condition.All metallic contacts on humid or wet condition create electrolytic reactions with contact corrosion as a result.Please find all reprocessing instructions for our products in the leaflet ¿important information¿.During the investigation, no product design or manufacturing issues were observed that may have contributed to the complaint condition.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.A conclusive determination could be reached and therefore the in the investigation flow listed remaining investigation steps are not required.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed as no product related issue could be detected device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TAP FOR 4.0MM CANCELLOUS BONE SCREWS/110MM
Type of Device
TAP,BONE
Manufacturer (Section D)
OBERDORF SYNTHES PRODUKTIONS GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
MDR Report Key9446372
MDR Text Key196991963
Report Number8030965-2019-70994
Device Sequence Number1
Product Code HWX
UDI-Device Identifier07611819019348
UDI-Public(01)07611819019348
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 11/22/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number311.34
Device Lot Number9894189
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/15/2020
Initial Date Manufacturer Received 11/22/2019
Initial Date FDA Received12/10/2019
Supplement Dates Manufacturer Received12/24/2019
02/07/2020
02/13/2020
09/15/2020
10/01/2020
Supplement Dates FDA Received01/17/2020
02/10/2020
02/13/2020
09/18/2020
10/06/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
TAP F/CANCELLOUSSCR Ø4 L110/85
-
-