DEPUY SYNTHES PRODUCTS LLC BATTERY OSCILLATOR II FOR BPL II; MOTOR, SURGICAL INSTRUMENT, AC-POWERED
|
Back to Search Results |
|
Catalog Number 530.710 |
Device Problem
Noise, Audible (3273)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 01/23/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Udi: (b)(4).Concomitant med products and therapy dates: blade device, (b)(6) 2019.Device evaluation: the actual device was returned for evaluation.A visual and functional assessment was performed on the device which determined that it failed for check saw blade coupling.During service/ repair, it was determined that the device made a grinding noise, the turn knob and straining ring had come loose from the saw head, and the oscillating head was worn.It was further determined that the issues were consistent with improper maintenance.Therefore, the reported condition was confirmed.The assignable root cause was determined to be traced to maintenance, which is improper maintenance.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Event Description
|
It was reported that during testing, it was observed that the battery oscillator device was not holding on to the blade device when attached.During in-house engineering evaluation, it was observed that the device made a grinding noise.The event was not related to surgery.There was no patient involvement reported.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred on (b)(6) in 2019.The actual month was not specified.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Search Alerts/Recalls
|
|
|