DEPUY SYNTHES PRODUCTS LLC BATTERY OSCILLATOR II FOR BPL II; MOTOR, SURGICAL INSTRUMENT, AC-POWERED Back to Search Results
Catalog Number 530.710
Device Problem Noise, Audible (3273)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/23/2019
Event Type  malfunction  
Manufacturer Narrative
Udi: (b)(4).Concomitant med products and therapy dates: blade device, (b)(6) 2019.Device evaluation: the actual device was returned for evaluation.A visual and functional assessment was performed on the device which determined that it failed for check saw blade coupling.During service/ repair, it was determined that the device made a grinding noise, the turn knob and straining ring had come loose from the saw head, and the oscillating head was worn.It was further determined that the issues were consistent with improper maintenance.Therefore, the reported condition was confirmed.The assignable root cause was determined to be traced to maintenance, which is improper maintenance.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Event Description
It was reported that during testing, it was observed that the battery oscillator device was not holding on to the blade device when attached.During in-house engineering evaluation, it was observed that the device made a grinding noise.The event was not related to surgery.There was no patient involvement reported.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred on (b)(6) in 2019.The actual month was not specified.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BATTERY OSCILLATOR II FOR BPL II
Type of Device
MOTOR, SURGICAL INSTRUMENT, AC-POWERED
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
OBERDORF SYNTHES PRODUKTIONS GMBH
eimattstrasse 3
oberdorf 4436
SZ   4436
Manufacturer Contact
kara ditty-bovard
eimattstrasse 3
oberdorf 4436
SZ   4436
6103142063
MDR Report Key9446858
MDR Text Key195161670
Report Number8030965-2019-70874
Device Sequence Number1
Product Code GEY
UDI-Device Identifier07611819491847
UDI-Public07611819491847
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial
Report Date 09/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/10/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number530.710
Device Lot Number103031
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/08/2019
Date Manufacturer Received11/28/2019
Date Device Manufactured08/09/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-