DEPUY SYNTHES PRODUCTS LLC BATTERY REAMER/DRILL FOR BPL II; MOTOR, SURGICAL INSTRUMENT, AC-POWERED
|
Back to Search Results |
|
Catalog Number 530.705 |
Device Problem
Physical Resistance/Sticking (4012)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 11/05/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Udi: (b)(4).Device evaluation: the actual device was returned for evaluation.During repair, it was determined that the reported condition was confirmed.The assignable root cause was determined to be traced to component failure (faulty parts), which is normal wear.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Event Description
|
It was reported by (b)(6) that during service and evaluation, it was determined that the moving parts of the trigger did not move smoothly on the battery reamer/drill device.It was further determined that the device failed pretest for check trigger and ecu (electric control unit) function and trigger test.It was noted in the service order that the device had an undetermined malfunction.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should available, a supplemental medwatch will be submitted accordingly.
|
|