• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL HUDSON NEBULIZER ADAPTOR 028,STERILE,SHELFPAK,; NEBULIZER (DIRECT PATIENT INTE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL HUDSON NEBULIZER ADAPTOR 028,STERILE,SHELFPAK,; NEBULIZER (DIRECT PATIENT INTE Back to Search Results
Catalog Number 031-28J
Device Problem Connection Problem (2900)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/22/2019
Event Type  malfunction  
Manufacturer Narrative
Qn# (b)(4).
 
Event Description
Customer reported the connection between the nut adaptor and the flow meter was unstable.A new unit was used.No patient harm or injury reported.
 
Manufacturer Narrative
(b)(4).The sample was received for analysis.A visual exam was performed and it was observed that the adaptor doesn't spin freely.No other issues were found.Oxygen entrainment testing was performed and during the setup it was observed that the assembly of the nut adaptor and the upper body was unstable.Even with that condition, the sample was able to be tested on oxygen entrainment testing but the testing failed due to the unstable condition.After the testing finished, the adaptor was carefully disassembled from the upper body and it was visually inspected.During the visual inspection wear was found on the internal tabs.Based on the investigation performed, the complaint is confirmed.Although the complaint is confirmed, there is not sufficient evidence to assure that this issue was originated during the manufacturing assembly or molding process.The wear on the internal tabs of the adaptor was most likely caused by the end user during the connection of the adaptor into the flowmeter that causes an unstable connection issue.
 
Event Description
Customer reported the connection between the nut adaptor and the flow meter was unstable.A new unit was used.No patient harm or injury reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUDSON NEBULIZER ADAPTOR 028,STERILE,SHELFPAK,
Type of Device
NEBULIZER (DIRECT PATIENT INTE
Manufacturer (Section D)
TELEFLEX MEDICAL
morrisville NC
MDR Report Key9447009
MDR Text Key191740219
Report Number3004365956-2019-00353
Device Sequence Number1
Product Code CAF
Combination Product (y/n)N
PMA/PMN Number
K153010
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 11/22/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/10/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Catalogue Number031-28J
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/30/2019
Date Manufacturer Received01/27/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
NONE REPORTED.; NONE REPORTED.
-
-