• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1 Back to Search Results
Model Number 816571
Device Problems Device Slipped (1584); Output Problem (3005)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/18/2019
Event Type  malfunction  
Manufacturer Narrative
The field service representative (fsr) was notified by the perfusionist that the issue was due to operator error.The fsr could not duplicate the error.
 
Event Description
It was reported that the pump displayed a belt slip error.No other details regarding the nature of this event were provided.
 
Manufacturer Narrative
Updated blocks: e1 and h6.The reported complaint was confirmed.Per data log analysis, on (b)(6) 2019, large roller pump module id 04904 reported many underspeed (head < demand) and an underspeed (belt slip).Looking through the log this has occurred on other days as well.The pump is labeled as vent in the perfusion screen.The log confirms the complaint.Per field service representative (fsr), the end user over occluded the roller pump which caused the errors.Further instruction was provided to the user on proper occlusion.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCED PERFUSION SYSTEM 1
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
MDR Report Key9451519
MDR Text Key199112437
Report Number1828100-2019-00652
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
PMA/PMN Number
K172220
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 02/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number816571
Device Catalogue Number816571
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/18/2019
Initial Date FDA Received12/11/2019
Supplement Dates Manufacturer Received01/29/2020
Supplement Dates FDA Received02/04/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-