Model Number X SERIES
Device Problem Pacing Problem (1439)
Patient Problem No Information (3190)
Event Type  malfunction  
Manufacturer Narrative
Zoll medical corporation evaluated the device and the customer's report was not replicated or confirmed.The device was put through extensive testing which included bench handling, stress testing, and full functionality testing with the customer's multifunction cable, cpr adapter, and a set of test pads without duplicating any device malfunction.No trend is associated with reports of this type.Review of the device logs did not find evidence to support the reported event.
 
Event Description
Complainant alleged that while attempting to pace a (b)(6) male patient, the device displayed an unspecified "pacer fault" message.Complainant did not indicate that there was any adverse effect to the patient due to the reported malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
X SERIES
Type of Device
DEFIBRILLATOR/PACEMAKER
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
Manufacturer Contact
269 mill road
chelmsford, MA 01824
9784219552
MDR Report Key9453579
MDR Text Key170449993
Report Number1220908-2019-03779
Device Sequence Number1
Product Code DRO
UDI-Device Identifier00847946005979
UDI-Public00847946005979
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112432/P160
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Other
Type of Report Initial
Report Date 11/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberX SERIES
Device Catalogue NumberX SERIES
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/25/2019
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/20/2019
Initial Date FDA Received12/11/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/01/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age70 YR
-
-