MAUDE Adverse Event Report: DEPUY MITEK LLC US VAPR TRIPOLAR 90 SUCTION ELECT; ENDOSCOPIC ELECTROSURGICAL ELECTRODE, BIPOLAR, SINGLE USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US VAPR TRIPOLAR 90 SUCTION ELECT; ENDOSCOPIC ELECTROSURGICAL ELECTRODE, BIPOLAR, SINGLE USE Back to Search Results
Catalog Number 225028
Device Problem Suction Failure (4039)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/26/2019
Event Type  malfunction  
Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.(b)(4).
 
Event Description
It was reported by affiliate that during an arthroscopic rotator cuff repair was found that the vapr tripolar 90 suction elect didn¿t suction.No patient consequence and no surgical delay was reported.No additional information was provided.The affiliate later reported the device would not be returning for evaluation.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Additional narrative: a manufacturing record evaluation was performed for the finished device lot/serial number u1906033, and no non-conformances were identified.The complaint device is not being returned, therefore unavailable for a physical evaluation.This complaint cannot be confirmed.A manufacturing record evaluation was performed for the finished device lot/serial number u1906033, and no non-conformances were identified.Since the complaint device was not returned, we cannot determine a root cause for the reported failure.If the device is received in the future, we will reopen the complaint and perform the investigation as appropriate.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VAPR TRIPOLAR 90 SUCTION ELECT
Type of Device
ENDOSCOPIC ELECTROSURGICAL ELECTRODE, BIPOLAR, SINGLE USE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key9454789
MDR Text Key208154517
Report Number1221934-2019-59938
Device Sequence Number1
Product Code GEI
UDI-Device Identifier10886705023103
UDI-Public10886705023103
Combination Product (y/n)N
PMA/PMN Number
K143475
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign
Type of Report Initial,Followup
Report Date 11/26/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2022
Device Catalogue Number225028
Device Lot NumberU1906033
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 11/26/2019
Initial Date FDA Received12/11/2019
Supplement Dates Manufacturer Received12/12/2019
Supplement Dates FDA Received12/23/2019
Patient Sequence Number1
-
-