• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES PRODUCTS LLC BATTERY OSCILLATOR II FOR BPL II; MOTOR, SURGICAL INSTRUMENT, AC-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES PRODUCTS LLC BATTERY OSCILLATOR II FOR BPL II; MOTOR, SURGICAL INSTRUMENT, AC-POWERED Back to Search Results
Catalog Number 530.710
Device Problem Noise, Audible (3273)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/01/2019
Event Type  malfunction  
Manufacturer Narrative
The actual device has been returned and is currently pending evaluation.Once the evaluation of the device has been completed, a supplemental medwatch report will be sent accordingly.Udi: (b)(4).
 
Event Description
It was reported from (b)(6) that during an unspecified surgical procedure it was observed that the battery oscillator device was ¿noisy and crunchy¿.It was not reported if there were any delays in the procedure due to the event.It was not reported if there was a spare device available for use.There was patient involvement reported.There were no reports of any injuries, medical intervention or prolonged hospitalization.The exact date of the event was not reported, however, it was reported that the event occurred in 2019.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Manufacturer Narrative
H10: depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Correction: d10: the date returned to manufacturer was documented as december 9, 2019 on the initial report.This date has been updated to december 13, 2019.If additional information should become available, a supplemental medwatch report will be submitted accordingly.Device evaluation: the actual device was returned for evaluation.Quality engineering evaluated the battery oscillator device and it was determined that the device had motor and bearing damage.Therefore, the reported condition that the device was noisy and crunchy was confirmed.It was further determined that the oscillating head was damaged (cracked and missing pins).It was further determined that the device failed pretest for check saw-blade coupling.It was determined that the root cause of the oscillation saw head damage/crack is a design related issue, and is covered under a capa.The assignable root cause of the motor and bearing damaged was determined to be traced to component failure due to normal wear.A review of the device history was performed and no non-conformances were detected related to the reported condition.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BATTERY OSCILLATOR II FOR BPL II
Type of Device
MOTOR, SURGICAL INSTRUMENT, AC-POWERED
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
eimattstrasse 3
oberdorf 4436
SZ  4436
MDR Report Key9459567
MDR Text Key193946840
Report Number8030965-2019-70997
Device Sequence Number1
Product Code GEY
UDI-Device Identifier07611819491847
UDI-Public7611819491847
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 11/27/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number530.710
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/13/2019
Initial Date Manufacturer Received 11/27/2019
Initial Date FDA Received12/12/2019
Supplement Dates Manufacturer Received01/16/2020
Supplement Dates FDA Received01/23/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-