DEPUY SYNTHES PRODUCTS LLC BATTERY OSCILLATOR II FOR BPL II; MOTOR, SURGICAL INSTRUMENT, AC-POWERED
|
Back to Search Results |
|
Catalog Number 530.710 |
Device Problem
Noise, Audible (3273)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 01/01/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
The actual device has been returned and is currently pending evaluation.Once the evaluation of the device has been completed, a supplemental medwatch report will be sent accordingly.Udi: (b)(4).
|
|
Event Description
|
It was reported from (b)(6) that during an unspecified surgical procedure it was observed that the battery oscillator device was ¿noisy and crunchy¿.It was not reported if there were any delays in the procedure due to the event.It was not reported if there was a spare device available for use.There was patient involvement reported.There were no reports of any injuries, medical intervention or prolonged hospitalization.The exact date of the event was not reported, however, it was reported that the event occurred in 2019.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Manufacturer Narrative
|
H10: depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Correction: d10: the date returned to manufacturer was documented as december 9, 2019 on the initial report.This date has been updated to december 13, 2019.If additional information should become available, a supplemental medwatch report will be submitted accordingly.Device evaluation: the actual device was returned for evaluation.Quality engineering evaluated the battery oscillator device and it was determined that the device had motor and bearing damage.Therefore, the reported condition that the device was noisy and crunchy was confirmed.It was further determined that the oscillating head was damaged (cracked and missing pins).It was further determined that the device failed pretest for check saw-blade coupling.It was determined that the root cause of the oscillation saw head damage/crack is a design related issue, and is covered under a capa.The assignable root cause of the motor and bearing damaged was determined to be traced to component failure due to normal wear.A review of the device history was performed and no non-conformances were detected related to the reported condition.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
|
Search Alerts/Recalls
|
|
|