MAUDE Adverse Event Report: MEDTRONIC CRYOCATH LP FLEXCATH ADVANCE STEERABLE SHEATH; CATHETER, STEERABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC CRYOCATH LP FLEXCATH ADVANCE STEERABLE SHEATH; CATHETER, STEERABLE Back to Search Results
Model Number 4FC12
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Chest Pain (1776); Myocardial Infarction (1969); ST Segment Elevation (2059)
Event Date 08/20/2019
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during a cryo ablation procedure, st segment elevation occurred after insertion of the balloon catheter and mapping catheter within the sheath.A coronary angiography was performed and coronary vasospasm was found.A myocardial infarction was indicated and the patient presented with chest pain.The case was aborted, and it was unknown whether the patient was under general anesthesia.The patient is a participant in the comparison of rf and cryoballoon ablation therapy of af clinical study.No further patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXCATH ADVANCE STEERABLE SHEATH
Type of Device
CATHETER, STEERABLE
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA  H9R 5Z8
Manufacturer (Section G)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA   H9R 5Z8
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key9459857
MDR Text Key170510364
Report Number3002648230-2019-00844
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
K123591
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 12/12/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/12/2019
Is this an Adverse Event Report? Yes
Device Operator Health Professional
Device Expiration Date03/28/2021
Device Model Number4FC12
Device Catalogue Number4FC12
Device Lot Number51127
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/25/2019
Date Device Manufactured03/29/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
2ACH20 MAPPING CATHETER
Patient Outcome(s) Hospitalization; Life Threatening; Required Intervention;
Patient Age54 YR
Patient Weight54
-
-