Catalog Number 530.610 |
Device Problem
Noise, Audible (3273)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 01/01/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Udi - (b)(4).The actual device was returned for evaluation.During repair, an evaluation was performed and it was determined that the reported condition was confirmed.The assignable root cause was determined to be due to improper maintenance.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
|
Event Description
|
It was reported from (b)(6) that during service and evaluation, it was determined that the battery oscillator device bearing and housing were worn, the device had cosmetic damage and the device made excessive noise.It was further determined that the device failed pretest for general condition.It was noted in the service order that the device was not working.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The date of event was unknown but was known to have occurred in 2019.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
|