The reported event was confirmed as manufacturing related.Visual evaluation of the returned sample noted one opened (without original packaging), silicone foley catheter with connected sample port connector and cut inlet tube portion.Visual inspection of the sample noted no obvious visible defects.The catheter balloon was attempted to be inflated with 10 ml methylene blue solution (3 drops 1% aq methylene blue per 100ml distilled water), and immediately leaked out of the funnel and eyelets, implying lumen to lumen leakage.The catheter balloon was dissected to find that the inflation notch perforates both lumens.This is out of specification per inspection procedure which states, "inflation lumen notch not to penetrate drainage lumen and must be properly formed, in addition no nicks are allowed." although the reported event was confirmed, the root cause could not be determined.A potential root cause for this failure could be, ¿incorrect set-up.¿ the device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "[warnings] 1.Method for use (1) do not inflate the balloon in the urethra.[the urethra may be injured.] (2) do not pull the catheter hard.[the bladder/urethra may be injuryed.] 2.Applicable patients patients with delirium who might pull out catheter [when patient tugs at catheter unconsciously, the bladder and urethra may be damaged.] [contraindications] 1.Method for use: (1) do not reuse (2) do not resterilize (3) this device contains 10% povidine-iodine.For patients with past history of allergic hypersensitivity to povidone-iodine or iodine, consider using alternative disinfectants.(4) be careful that the catheter is not exposed to ointments, contract medium or oil-based lubricants (including vegetable oils such as olive oil, mineral oils such as white petrolatum and animal oils).[they may damage the device and may burst balloon.] (5) do not hold the device with forceps, etc.Avoid contact with any blades or sharp-edged instruments.[catheter damage may cause balloon rupture and accidental balloon removal or failure to deflate or remove the balloon.] directions for use: 1) clean the area around the external urethral meatus with the packaged cotton balls immersed in the antiseptics.2) lubricate the catheter shaft with the lubricant jelly.3) carefully insert the catheter into the urethral meatus.After the balloon advanced in the bladder, attach the needleless syringe, and gently infuse the specified volume of sterile water to inflate the balloon.4) pull the catheter slightly to seat the balloon at the level of the bladder neck.5) to deflate and remove the balloon, attach a needleless syringe to let sterile water in the balloon come out spontaneously through balloon deflation without aspiration.After balloon deflation, withdraw the catheter slowly while confirming that no abnormal resistance is encountered." h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|