Catalog Number EVX35-06-060-080 |
Device Problems
Migration or Expulsion of Device (1395); Device Dislodged or Dislocated (2923)
|
Patient Problems
Unspecified Infection (1930); Pain (1994); Perforation of Vessels (2135)
|
Event Date 12/03/2019 |
Event Type
Injury
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Physician used an everflex entrust with a non-medtronic 6fr sheath and 190cm guidewire during treatment of the patient¿s mid right superficial femoral artery (sfa).Embolic protection was not used.Procedure completed as per ifu without issue.The patient complained of leg pain.Follow up exams performed showed the stent was in an extra-vascular space.Explant in or performed which showed infection.The stent was kept for pathologic analysis, of which the preliminary results showed gram negative staphylococcus epidermis, gram positive staphylococcus eperdermis, and pseudomonas aeruginosa.It is also reported that vessel damage/injury/perforation occurred.
|
|
Manufacturer Narrative
|
Additional information: the surgeon believes the perforation was caused by tissue necrosis from the infection.No further treatment was required.Patient has been discharged from the hospital and is doing fine.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Additional information: stent implant date provided.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|