• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO PHILIPPINES CORPORATION MCKESSON MEDI-PAK PERFORMANCE SAFETY HYPODERMIC NEEDLE; SYRINGE, ANTISTICK

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO PHILIPPINES CORPORATION MCKESSON MEDI-PAK PERFORMANCE SAFETY HYPODERMIC NEEDLE; SYRINGE, ANTISTICK Back to Search Results
Model Number N/A
Device Problem Activation Problem (4042)
Patient Problems No Consequences Or Impact To Patient (2199); Needle Stick/Puncture (2462)
Event Date 10/11/2019
Event Type  malfunction  
Manufacturer Narrative
The actual device was not returned to the manufacturing facility for evaluation.Based from the results of our investigation, the root cause of the complaint could not be identified.The condition of the actual sample was not confirmed since it was not returned for evaluation.Retention samples were confirmed free from defects that will affect activation of safety sheath.Related functional testing such as sheath radial strength, sheath activation and deactivation were all passed.In addition, the safety sheath was successfully activated without any difficulty during simulation.We have series of visual in-process inspection to detect abnormality on the sheath that may lead to problem during sheath activation.Molding condition of the components critical to the safety activation of the product is routinely checked to assure that no defects will be encountered that will lead to sheath activation problem.Similarly, the assembly status of the safety needle such as sheath collar fitting and damaged parts that will affect product function is being confirmed.Lot history file revealed no related nonconformity or irregularity that will lead to the complaint.Prior shipment, qc conducts outgoing visual, sensory, and functional inspection to assure lots are in good quality.All samples passed.(b)(4).
 
Event Description
The user facility reported that an employee that incurred a needle stick because the mckesson brand safety needle was difficult to close.The green safety hinge that covers the needle usually is flipped and closed while pushing against the patient table, as to dispose of with needle covered to be safe.However, the cover is not closing easily and flips back up with the needle still out, sticking the nurse.Additional information was received 19novermber2019: no nurses sought medical treatment.There were no risks of infectious disease.The patients past medical history indicates any potential infectious disease which a nurse would be exposed to by blood bourne pathogens.None of the pediatric patients had any infectious diseases.The nurses was stuck when the needle poked them through the finger or into their cuticle.The dirty needles go into a biohazard container which is then incinerated by a waste management company.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MCKESSON MEDI-PAK PERFORMANCE SAFETY HYPODERMIC NEEDLE
Type of Device
SYRINGE, ANTISTICK
Manufacturer (Section D)
TERUMO PHILIPPINES CORPORATION
124 east main avenue
laguna technopark
binan,
RP 
Manufacturer (Section G)
TERUMO PHILIPPINES CORPORATION
124 east main avenue
laguna technopark
binan, laguna,,
RP  
Manufacturer Contact
theresa mussaw
950 elkton blvd.
elkton, MD 21921
8002837866
MDR Report Key9474269
MDR Text Key219786477
Report Number3003902955-2019-00052
Device Sequence Number1
Product Code MEG
UDI-Device Identifier30612479178763
UDI-Public30612479178763
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K051865
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,user facility
Reporter Occupation Other
Type of Report Initial
Report Date 12/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/16/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2024
Device Model NumberN/A
Device Catalogue Number102-N231S
Device Lot Number190527A
Was Device Available for Evaluation? No
Date Manufacturer Received11/18/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/27/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-