• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION WATCHMAN ACCESS SYSTEM; SYSTEM, APPENDAGE CLOSURE, LEFT ATRIAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION WATCHMAN ACCESS SYSTEM; SYSTEM, APPENDAGE CLOSURE, LEFT ATRIAL Back to Search Results
Model Number 10374
Device Problem Material Deformation (2976)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/03/2019
Event Type  malfunction  
Event Description
It was reported that a device kink occurred.A left atrial appendage (laa) closure procedure was being performed.A watchman access system (was) was positioned and a watchman laa closure device & delivery system (wds) were used.The was was positioned in the laa and the wds was inserted into it.The closure device was deployed in the laa, but did not meet release criteria and needed to be fully recaptured.After this device was removed from the patient the was was noted to have kinked.The procedure was continued after this was noted and completed, there were no other complications that occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WATCHMAN ACCESS SYSTEM
Type of Device
SYSTEM, APPENDAGE CLOSURE, LEFT ATRIAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key9477180
MDR Text Key176631870
Report Number2134265-2019-15638
Device Sequence Number1
Product Code NGV
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
P130013
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 12/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/20/2022
Device Model Number10374
Device Catalogue Number10374
Device Lot Number0023383540
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/05/2019
Initial Date FDA Received12/16/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/21/2019
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Age66 YR
Patient Weight80
-
-