• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. LIGACLIP*ENDO ROTATING MCA; CLIP, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. LIGACLIP*ENDO ROTATING MCA; CLIP, IMPLANTABLE Back to Search Results
Model Number ER320
Device Problem Material Split, Cut or Torn (4008)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/25/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Batch # unk.A manufacturing record evaluation was performed for the finished device lot number and no non-conformances were identified.
 
Event Description
It was reported that during an unknown procedure, there was a tear in package.A like device was used to complete the procedure with no patient consequences reported.
 
Manufacturer Narrative
(b)(4).Date sent: 1/10/2020 d4: batch # t40u71 investigation summary the analysis results found that the er320 device was returned inside its package opened.Upon visual inspection, it was observed that the blister from the packaging was damaged.It was noted to be broken and still adhered to the tyvek.Due to the damages found on the blister, a possible cause for this condition is due to improper handling during transit or storage.It appears that the package hit a hard surface and this caused the reported event.The reported complaint was confirmed.All ethicon product is 100% inspected prior to release.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
MDR Report Key9478661
MDR Text Key203868684
Report Number3005075853-2019-24427
Device Sequence Number1
Product Code FZP
UDI-Device Identifier10705036012580
UDI-Public10705036012580
Combination Product (y/n)N
PMA/PMN Number
K864102
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 11/26/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberER320
Device Catalogue NumberER320
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/13/2019
Initial Date Manufacturer Received 11/26/2019
Initial Date FDA Received12/17/2019
Supplement Dates Manufacturer Received12/13/2019
Supplement Dates FDA Received01/10/2020
Patient Sequence Number1
-
-