• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ASPEN SURGICAL PRODUCTS, CALEDONIA RICHARD-ALLAN REGULAR TIP PEN W/ RULER AND 6-UP LABELS; MARKING PEN

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ASPEN SURGICAL PRODUCTS, CALEDONIA RICHARD-ALLAN REGULAR TIP PEN W/ RULER AND 6-UP LABELS; MARKING PEN Back to Search Results
Model Number 2650
Device Problem Defective Device (2588)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/03/2019
Event Type  malfunction  
Manufacturer Narrative
No further information is available on the product at this time.The investigation is ongoing, however if any additional relevant information is identified following completion of the investigation, the additional relevant information will be submitted in a supplemental report.Device not returned.
 
Event Description
Aspen surgical received a report from the distributor indicating that richard-allan regular tip pen was found with defective seals.The item was not in use.No injury/death was reported.This report was filed in our complaint handling system as number (b)(4).
 
Manufacturer Narrative
Aspen surgical received a report from the distributor indicating that product was found with seal issues.The actual device was returned for evaluation.The manufacturing lot number along with samples were provided for review.The distributor indicated that the defects were found during incoming inspection.A review of the samples confirmed the issue from the distributor.Analysis of the finished good lot number was reviewed.No non-conformance's were noted during the manufacturing process.According to the manufacturing process, these parts are loaded into recessed pockets on a packaging machine per specified work instructions.If incorrectly loaded or not orientated properly, the product interferes with the sealing process.The ifu which is received with the product, along with the pouch label, identifies this failure mode with the symbol "do not use if package is damaged".This indicates that the device should not be used if the products sterile barrier system or its packaging is compromised.Additionally, production supervisors and operations were notified of this issue.Based on this information, no further action is required.
 
Event Description
Aspen surgical received a report from the distributor indicating that richard-allan regular tip pen was found with defective seals.The item was not in use.No injury/death was reported.This report was filed in our complaint handling system as number (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RICHARD-ALLAN REGULAR TIP PEN W/ RULER AND 6-UP LABELS
Type of Device
MARKING PEN
Manufacturer (Section D)
ASPEN SURGICAL PRODUCTS, CALEDONIA
6945 southbelt dr. s.e.
caledonia MI 49316
MDR Report Key9480434
MDR Text Key191438147
Report Number1836161-2019-00101
Device Sequence Number1
Product Code FZZ
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Type of Report Initial,Followup
Report Date 12/03/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number2650
Device Lot Number196686
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/03/2019
Initial Date FDA Received12/17/2019
Supplement Dates Manufacturer Received12/03/2019
Supplement Dates FDA Received02/17/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-