This follow-up report is being submitted to relay additional information. visual evaluation of the returned products identified that the stem has protruded through inner and outer packaging damaging both sterile cavities and the outer carton box.Sterility has been compromised.Complaint sample was evaluated and the reported event was confirmed.Device history record was reviewed and no discrepancies were found.This product was likely conforming when it left zimmer biomet control.The root cause of the reported issue is attributed to transit damage.This device falls within the scope of a corrective action which is to assess all current sterile barrier systems used to package products at zimmer biomet bridgend.As part of this corrective action, the pouch is being improved to use a stronger material (nylon), and foam end caps are being added.Also, the orientation the devices are packed in the shipper box is moving from vertical to horizontal, and the thickness of the shipper box has been increased.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|