ZIMMER BIOMET, INC. TPRLC 133 FP 12/14 BM HO 6.0; PROSTHEIS, HIP Back to Search Results
Model Number N/A
Device Problem Packaging Problem (3007)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/21/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that during an initial hip arthroplasty , the sterile package was damaged.No adverse events have been reported as a result of the malfunction, attempts have been made and additional information on the reported event is unavailable.
 
Event Description
No further event information available at the time of this report.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.  visual evaluation of the returned products identified that the stem has protruded through inner and outer packaging damaging both sterile cavities and the outer carton box.Sterility has been compromised.Complaint sample was evaluated and the reported event was confirmed.Device history record was reviewed and no discrepancies were found.This product was likely conforming when it left zimmer biomet control.The root cause of the reported issue is attributed to transit damage.This device falls within the scope of a corrective action which is to assess all current sterile barrier systems used to package products at zimmer biomet bridgend.As part of this corrective action, the pouch is being improved to use a stronger material (nylon), and foam end caps are being added.Also, the orientation the devices are packed in the shipper box is moving from vertical to horizontal, and the thickness of the shipper box has been increased.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TPRLC 133 FP 12/14 BM HO 6.0
Type of Device
PROSTHEIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key9480802
MDR Text Key196447375
Report Number0001825034-2019-05598
Device Sequence Number1
Product Code KWY
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 04/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/17/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number51-131060
Device Lot Number3450091
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/19/2019
Was the Report Sent to FDA? No
Date Manufacturer Received04/14/2020
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
-
-