• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL HEMOLOK ML CLIPS 6/CART 84/BOX; CLIP, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL HEMOLOK ML CLIPS 6/CART 84/BOX; CLIP, IMPLANTABLE Back to Search Results
Catalog Number 544230
Device Problem Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/26/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device history review for the product hemolok ml clips 6/cart 84/box lot# 73d1900257 investigation did not show issues related to the complaint.The device has not been returned for investigation.Teleflex will continue to monitor and trend related events.
 
Event Description
It was reported that a clip was found broken at opening the cartrige and could not be loaded to applier.Therefore, a new cartridge was opened instead.
 
Event Description
It was reported that a clip was found broken at opening the cartrige and could not be loaded to applier.Therefore, a new cartridge was opened instead.
 
Manufacturer Narrative
Qn#(b)(4).The customer returned one cartridge 544230 hemolok ml clips 6/cart 84/box for investigation.The cartridge was visually examined with and without magnification.Visual examination of the cartridge revealed that the cartridge was returned with a broken clip and no intact clips remaining in it.The broken clip was removed from the cartridge and it was observed that it was broken in half at the hinge.The clip breaking at the hinge during loading was determined to be the result of inadequate cross-sectional area at the outer hinge.The root cause is due to inadequate specification of hinge thickness, which is a design related issue.A capa has been previously opened to further investigate issues related to clip breakages.Reference file anp1900073862 for investigation photos.The ifu for this product, l06110, was reviewed as a part of this complaint investigation.The ifu states, "always check the alignment of the applier jaws before use.When closed, jaw tips should be directly aligned and not offset.Alignment of the jaw is critical for safe application of the clip.If this is not done, patient injury may occur.Proper maintenance, care and cleaning are necessary to ensure proper functionality." the returned clip was broken in half at the hinge.The root cause is due to inadequate specification of hinge thickness, which is a design related issue.A capa has been previously opened to further investigate issues related to clip breakages.The reported complaint of "broken/detached parts - clip - hinge" was confirmed based upon the sample received.One cartridge was returned with a broken clip.The returned clip was broken in half at the hinge.The clip breaking at the hinge during loading was determined to be the result of inadequate cross-sectional area at the outer hinge.The root cause is due to inadequate specification of hinge thickness, which is a design related issue.A capa has been previously opened to further investigate issues related to clip breakages.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEMOLOK ML CLIPS 6/CART 84/BOX
Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
TELEFLEX MEDICAL
morrisville NC
MDR Report Key9482063
MDR Text Key173224272
Report Number3003898360-2019-01504
Device Sequence Number1
Product Code FZP
Combination Product (y/n)N
PMA/PMN Number
K133202
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial,Followup
Report Date 12/02/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/08/2024
Device Catalogue Number544230
Device Lot Number73D1900257
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/13/2020
Initial Date Manufacturer Received 12/02/2019
Initial Date FDA Received12/17/2019
Supplement Dates Manufacturer Received01/21/2020
Supplement Dates FDA Received01/27/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-