Qn#(b)(4).The customer returned one cartridge 544230 hemolok ml clips 6/cart 84/box for investigation.The cartridge was visually examined with and without magnification.Visual examination of the cartridge revealed that the cartridge was returned with a broken clip and no intact clips remaining in it.The broken clip was removed from the cartridge and it was observed that it was broken in half at the hinge.The clip breaking at the hinge during loading was determined to be the result of inadequate cross-sectional area at the outer hinge.The root cause is due to inadequate specification of hinge thickness, which is a design related issue.A capa has been previously opened to further investigate issues related to clip breakages.Reference file anp1900073862 for investigation photos.The ifu for this product, l06110, was reviewed as a part of this complaint investigation.The ifu states, "always check the alignment of the applier jaws before use.When closed, jaw tips should be directly aligned and not offset.Alignment of the jaw is critical for safe application of the clip.If this is not done, patient injury may occur.Proper maintenance, care and cleaning are necessary to ensure proper functionality." the returned clip was broken in half at the hinge.The root cause is due to inadequate specification of hinge thickness, which is a design related issue.A capa has been previously opened to further investigate issues related to clip breakages.The reported complaint of "broken/detached parts - clip - hinge" was confirmed based upon the sample received.One cartridge was returned with a broken clip.The returned clip was broken in half at the hinge.The clip breaking at the hinge during loading was determined to be the result of inadequate cross-sectional area at the outer hinge.The root cause is due to inadequate specification of hinge thickness, which is a design related issue.A capa has been previously opened to further investigate issues related to clip breakages.
|