BOSTON SCIENTIFIC CORPORATION ULTRAFLEX TRACHEOBRONCHIAL; PROSTHESIS, TRACHEAL, EXPANDABLE Back to Search Results
Model Number M00564770
Device Problems Positioning Problem (3009); Activation Failure (3270)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/26/2019
Event Type  Injury  
Manufacturer Narrative
Initial reporter's address 1: (b)(6).(b)(4).The device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
 
Event Description
It was reported to boston scientific corporation on (b)(6) 2019 that an ultraflex tracheobronchial covered distal release stent was to be used to treat a broncho-pleural fistula in the right main bronchus during a tracheal stent implantation procedure performed on (b)(6) 2019.Reportedly, patient's anatomy was dilated prior to stent placement.According to the complainant, during the procedure, the stent was able to be deployed.However, it was noticed that the stent failed to expand.The physician used a balloon catheter to dilate the stent but the stent could not be fully expanded, and the stent sipped towards the distal end of the bronchus.The stent was removed with forceps and another ultraflex tracheobronchial stent was placed to complete the procedure.There were no patient complications reported as a result of this event.The patient's condition following the procedure was reported to be stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRAFLEX TRACHEOBRONCHIAL
Type of Device
PROSTHESIS, TRACHEAL, EXPANDABLE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key9482618
MDR Text Key175235192
Report Number3005099803-2019-05986
Device Sequence Number1
Product Code JCT
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K012883
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 12/17/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/05/2020
Device Model NumberM00564770
Device Catalogue Number6477
Device Lot Number0021825075
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/26/2019
Initial Date FDA Received12/17/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/06/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age58 YR
Patient Weight76
-
-