The reported issue was confirmed.The device was returned for evaluation.Visual inspection found a pinhole on the rubberized layer, which caused water to leak out through the drainage eye.The reported event was found as a manufacturing related issue.The device history record was reviewed and found a possible manufacturing issue(s) that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: ¿warning: do not use ointments or lubricants having a petrolatum base.They will damage latex.Visually inspect the product for any imperfections or surface deterioration prior to use.¿ use luer tip syringe to inflate with stated ml of sterile water.Or ¿ for pre-filled products, remove clip and squeeze reservoir to inflate with stated ml of sterile water" h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|