Model Number 5392 |
Device Problem
Connection Problem (2900)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 12/06/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Product analysis:analysis could not confirm the customer comment.Output connectors were not broken.Could not fully insert the pacing cable.Atrial output connector was contaminated inside and out.Hanger was cracked.Keypad window was scratched.It was contaminated under rate and menu knobs.Both output nuts were contaminated.All found defective parts were replaced and all other identified issues were resolved.The device then passed all final functional tests.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that the external pulse generator (epg) had broken atrial and ventricular ports.The caller was advised to return the epg to service; the epg remains in use at this time.No patient complications have been reported as a result of this event.The epg was returned for service.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that the external pulse generator (epg) had broken atrial and ventricular ports.The caller was advised to return the programmer to service; the programmer remains in use at this time.No patient complications have been reported as a result of this event.
|
|