• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET UK LTD. OXF ANAT BRG LT MD SIZE 4 PMA; KNEE PROSTHESIS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET UK LTD. OXF ANAT BRG LT MD SIZE 4 PMA; KNEE PROSTHESIS Back to Search Results
Model Number N/A
Device Problem Insufficient Information (3190)
Patient Problem Pain (1994)
Event Date 11/26/2019
Event Type  Injury  
Manufacturer Narrative
(b)(4).Concomitant medical products: medical product: oxf twin-peg cmntd fem md pma catalog #: 161469 lot #: 236750, medical product: oxf uni tib tray sz e lm pma catalog #: 154726 lot #: 184970.Multiple mdr reports were filed for this event, please see associated reports: 3002806535-2019-00931, 3002806535-2019-00933.The investigation is in process.Once the investigation is complete, a follow-up mdr will be submitted.
 
Event Description
It was reported that a patient underwent an initial left knee arthroplasty.Subsequently, a revision procedure due to pain.
 
Manufacturer Narrative
(b)(4).This final report is being submitted to relay additional information.The following sections were updated: b4, b5, d4, g4, g7, h1, h2, h4, h6, h10.D10: the product has not been returned to zimmer biomet for investigation.D11: medical product: oxf twin-peg cmntd fem md pma, catalog #: 161469, lot #: 236750.Medical product: oxf uni tib tray sz e lm pma, catalog #: 154726, lot #: 184970.Multiple mdr reports were filed for this event, please see associated reports: 3002806535-2019-00933-1, 3002806535-2019-00931-1.As the product has not been received, the investigation was limited to the information provided; a review of device history records and complaint history.In addition, we have not been provided with x-rays or any supporting documentation which could provide additional information.A review of the manufacturing history records confirms no abnormalities or deviations reported.A review of the complaint database over the last 3 years has found 11 similar complaints for this item code 161469, no similar complaint for the item code 154726 and 2 similar complaints for this item code 159548.Regulatory assessment determines that the occurrence of the event and the risk remains within the acceptable limits identified in the risk management report.There were trends identified from the complaint history review.Similar hospitals can be found for several complaints.There are no trends in lot number and cause.In most cases of similar complaints, the cause could not be established.The root cause of the issue could not be determined with the information currently available, therefore the specific failure cause within the risk table could not be selected for comparison.
 
Event Description
It was reported that a patient underwent an initial left knee arthroplasty.Subsequently, a revision procedure due to pain was performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OXF ANAT BRG LT MD SIZE 4 PMA
Type of Device
KNEE PROSTHESIS
Manufacturer (Section D)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK  CF31 3XA
MDR Report Key9488453
MDR Text Key174207474
Report Number3002806535-2019-00930
Device Sequence Number1
Product Code NRA
Combination Product (y/n)N
PMA/PMN Number
P010014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 02/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date02/24/2022
Device Model NumberN/A
Device Catalogue Number159548
Device Lot Number707850
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/27/2019
Initial Date FDA Received12/18/2019
Supplement Dates Manufacturer Received02/24/2020
Supplement Dates FDA Received02/29/2020
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-