ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE) QUARTET; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 1458Q/86
Device Problems Failure to Capture (1081); Over-Sensing (1438)
Patient Problems Syncope (1610); Arrhythmia (1721); Chest Pain (1776); Dizziness (2194)
Event Type  Injury  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.Further information was requested but not received.
 
Event Description
It was reported that when the symptomatic patient presented in clinic for a follow up, during device interrogation, loss of capture with high ventricular rate were observed on the left ventricular lead.The patient was passed out for a few seconds and felt heart racing with dizziness.The patient also has been experienced intermittent chest pain in the past few months.The patient was transferred to the hospital.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer Contact
elizabeth boltz
15900 valley view court
sylmar, CA 91342
MDR Report Key9496663
MDR Text Key172040131
Report Number2938836-2019-17052
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734503198
UDI-Public05414734503198
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 12/19/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/19/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2021
Device Model Number1458Q/86
Device Lot NumberA000072571
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received12/03/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/25/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Weight104
-
-