H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the device was returned to the service facility for evaluation.During evaluation, the reported issue of scanner is not holding a charge/abrupt shut off was confirmed.The scanner was inspected and fully charged the unit shuts down prematurely when ac power is removed, the cause of the issue is due to an internal li-ion battery failure.The device was serviced, tested and returned to the customer.A history review of serial number (b)(6) showed no other similar product complaint(s) from this serial number.
|