H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the device was returned to the service facility for evaluation.During evaluation, the reported issue of screen is only showing half of the monitor was unconfirmed.The root cause of the reported failure is inconclusive as the reported issue could not be reproduced during evaluation.The scanner was inspected and operating normally and showing everything, the ultrasound screen is scanning normally.The device was serviced, tested returned to the customer.A history review of serial number (b)(6) showed no other similar product complaint(s) from this serial number.
|