DEPUY SYNTHES PRODUCTS LLC BATTERY REAMER/DRILL FOR BPL II; MOTOR, SURGICAL INSTRUMENT, AC-POWERED
|
Back to Search Results |
|
Catalog Number 530.705 |
Device Problem
Physical Resistance/Sticking (4012)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 01/01/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Udi: (b)(4).The actual device was returned for evaluation.During repair, an evaluation was performed and it was determined that the device had a sticky trigger and failed pretest for trigger test.It was further noted that the trigger was jammed.Therefore, the reported condition was confirmed.The assignable root cause was determined to be traced to component failure which is wear from normal use and servicing over time.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
|
Event Description
|
It was reported from (b)(6) that the trigger button of the battery reamer/drill device was stuck.It was not reported if the device was used in surgery, or if there was patient involvement.It was not reported if there were any delays in a surgical procedure or if a spare device was available.It was not reported if there were any injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in 2019.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Search Alerts/Recalls
|
|
|