The customer contacted a siemens customer care center (ccc) and reported that discordant, falsely low complete blood count (cbc) parameters were obtained on two patient samples on an advia 2120i hematology system with dual aspirate autosampler.A siemens customer service engineer (cse) was dispatched to the customer site.The cse found the reagent leaking from the reagent tubing which was twisted and split.The tubing was replaced, the reagent block was cleaned, the analyzer was cleaned and the reagents in the system were primed.In a follow-up visit, the cse ran priming for the reagents and ran several precision runs.No outliers were observed during the runs after the reagent lines and pump were replaced.Quality controls (qc) were run and recovered within specifications.The cause of the discordant results was a combination of the damaged reagent tubing and the dirty reagent block.The system is performing according to specifications.No further evaluation of this device is required.
|
Discordant, falsely low complete blood count (cbc) parameters were obtained on two patient samples on an advia 2120i hematology system with dual aspirate autosampler.The discordant parameters include white blood cells (wbc), red blood cells (rbc), hemoglobin (hgb), hematocrit (hct), and platelets (plt).The discordant results were not reported to the physician(s).The two patient samples were repeated for wbc, rbc, hgb, hct, and plt on an alternate advia 2120i hematology system, resulting higher.These results were reported, as the correct results, to the physician(s).There are no reports of patient intervention or adverse health consequences due to the discordant, falsely low white blood cells (wbc), red blood cells (rbc), hemoglobin (hgb), hematocrit (hct), and platelets (plt) results.
|