Model Number 72404251
Device Problems Inflation Problem (1310); Mechanical Problem (1384)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/07/2019
Event Type  Injury  
Manufacturer Narrative
Reservoir model- 72404155, lot sn- (b)(4), mfg date- 10/08/2012, exp date- 09/24/2014, gtin- (b)(4).Device not returned for evaluation.
 
Event Description
It was reported that due to an unspecified reason the patient had his inflatable penile prosthesis lgx (ipp) removed and replaced.Additional information was received that this surgery occurred due to an "inoperable ipp" because the implant would not inflate and broken tubing.The patient is now doing fine according to the sales representative.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMS INFLATABLE PENILE PROSTHESIS WITH INHIBIZONE
Type of Device
DEVICE IMPOTENCE MECHANICAL/HYDRAULIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
10700 bren road w
minnetonka MN 55343
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
10700 bren road west
minnetonka MN 55343
Manufacturer Contact
alyson harris
10700 bren road w
minnetonka, MN 55343
4089353452
MDR Report Key9500876
MDR Text Key172410831
Report Number2183959-2019-68277
Device Sequence Number1
Product Code FHW
UDI-Device Identifier00878953003597
UDI-Public00878953003597
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
N970012
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Physician
Type of Report Initial
Report Date 12/19/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date05/01/2014
Device Model Number72404251
Device Catalogue Number72404251
Device Lot Number770878001
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/03/2019
Initial Date FDA Received12/20/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/14/2012
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age70 YR
-
-