The investigation determined that a non-reproducible, lower than expected creatinine (crea) result was obtained from a non-vitros (randox) quality control (qc) fluid when tested using vitros chemistry products crea slides lot 1519-3490-8161 on a vitros 5600 integrated system.The assignable cause is likely related to the single microslide, although the issue with the specific slide could not be determined.A review of e-connectivity was performed by an ortho second level engineer and it was determined that all of the other slides in the same cartridge resulted as expected, indicating an issue with the cartridge is not likely.Additionally, the qc fluid is not a likely cause of the event as all other assays that used fluid from the same cup all resulted as expected.The microslide incubator slot that contained the slide that produced the lower than expected result was also reviewed, and it was determined that other assays were resulting as expected form the same slot; therefore, an issue with that particular slot is unlikely.The signal read for the slide in question was abnormally low compared to what is normally expected for crea at the concentration of the l2 qc fluid as mentioned in the complaint investigation, which further supports an issue with the single slide.Historical quality control results were provided for vitros crea reagent lot 1519-3490-8161; however the baseline mean and sd was not provided for comparison, therefore a definitive determination cannot be made in regards to performance.The qc appears to be precise when comparing to the ortho performance verifier assay sheet sd¿s at a similar concentration.Additionally, there are not any out of range sd flags continually posting from the historical qc from e-connectivity, indicating that the qc was likely acceptable.Continual tracking and trending of complaint data has not identified any signals to suggest there is a systemic quality issue with vitros crea reagent lot 1519-3490-8161.A vitros crea within run precision test was not performed, however an issue with the vitros 5600 integrated system is unlikely as all other assays were performing as expected, in addition to all other vitros crea results resulting as expected.
|
A customer obtained a non-reproducible, lower than expected creatinine (crea) result from a non-vitros (randox) quality control (qc) fluid when tested using vitros chemistry products crea slides lot 1519-3490-8161 on a vitros 5600 integrated system.Randox l2 qc result <13.260 umol/l vs.The expected result of 184.0 umol/l.Biased results of the direction and magnitude observed may lead to inappropriate physician action if undetected.The lower than expected vitros crea result occurred on a non-vitros control fluid.There has been no allegations of actual patient harm as a result of this event, however the investigation cannot conclude that patient sample results had not been affected or would not be affected if the event were to recur undetected.This report corresponds to ortho clinical diagnostics inc (ortho).Complaint numbers (b)(4).
|