Catalog Number 225028
Device Problem Loss of or Failure to Bond (1068)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/05/2019
Event Type  malfunction  
Manufacturer Narrative
Device was used for treatment, not diagnosis.Udi: (b)(4).A manufacturing record evaluation was performed for the finished device u1906035 number, and no non-conformances were identified.As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.
 
Event Description
It was reported by affiliate in (b)(6) that during a synovectomy treating shoulder joint contracture, it was observed that the clotting feature of a vapr tripolar90 suction electrode device did not work.There was no delay in the surgical procedure.There was a spare device available for use to complete the surgery.There was patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: investigation summary: the complaint device is not being returned, therefore unavailable for a physical evaluation.This complaint cannot be confirmed.With the information provided, and without the complaint device to evaluate, we cannot determine a root cause for the reported failure.A manufacturing record evaluation was performed for the finished device lot number [u1906035], and no non-conformances were identified.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Device history lot a manufacturing record evaluation was performed for the finished device lot number [u1906035], and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VAPR TRIPOLAR 90 SUCTION ELECT
Type of Device
ENDOSCOPIC ELECTROSURGICAL ELECTRODE, BIPOLAR, SINGLE USE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key9505650
MDR Text Key206004353
Report Number1221934-2019-60090
Device Sequence Number1
Product Code GEI
UDI-Device Identifier10886705023103
UDI-Public10886705023103
Combination Product (y/n)N
PMA/PMN Number
K143475
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 12/06/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2022
Device Catalogue Number225028
Device Lot NumberU1906035
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/06/2019
Initial Date FDA Received12/20/2019
Supplement Dates Manufacturer Received01/02/2020
Supplement Dates FDA Received01/03/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-