• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. LIGAMAX-5MM ENDO CLIP APPLIER; CLIP, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. LIGAMAX-5MM ENDO CLIP APPLIER; CLIP, IMPLANTABLE Back to Search Results
Model Number EL5ML
Device Problem Failure to Form Staple (2579)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/04/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Batch # unk.The lot/batch was not provided; therefore, a manufacturing record evaluation could not be performed.Attempts are being made to retrieve the device.To date, the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Event Description
It was reported that during a laparoscopic colectomy, the surgeon felt something unusual when the clip was fed into the jaws for the first firing, and a tear drop-shaped clip was formed while feeding the clip.The device was used on the blood vessel.After that, the surgeon tried to feed a clip into the jaws outside the patient for functionality, but the issue was not resolved.Another device was used to complete the case.There were no adverse consequences to the patient.
 
Manufacturer Narrative
(b)(4).Date sent: 1/17/2020.D4: batch # t94u28.Investigation summary: the analysis results found that the el5ml device was returned with no damage in the external components.In an attempt to replicate the reported incident, the instrument was tested for functionality.During the analysis, the device was cycled and it fed and formed 7 conforming clips.The event described could not be confirmed as the device performed without any difficulties noted.The reported complaint could not be confirmed.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGAMAX-5MM ENDO CLIP APPLIER
Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
MDR Report Key9506817
MDR Text Key191870390
Report Number3005075853-2019-24687
Device Sequence Number1
Product Code FZP
UDI-Device Identifier10705036001843
UDI-Public10705036001843
Combination Product (y/n)N
PMA/PMN Number
K050344
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 12/04/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEL5ML
Device Catalogue NumberEL5ML
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/06/2020
Initial Date Manufacturer Received 12/04/2019
Initial Date FDA Received12/20/2019
Supplement Dates Manufacturer Received01/06/2020
Supplement Dates FDA Received01/17/2020
Patient Sequence Number1
-
-