• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NEUROSURGERY STRATA II VALVE, REGULAR; SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NEUROSURGERY STRATA II VALVE, REGULAR; SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS Back to Search Results
Model Number 42866
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information received reported there were no environmental/external/patient factors that may have led or contributed to the issue.Temporarily, there are no actions/interventions taken to resolve the issue; however, the subsequent plan was to change to a va shunt.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported after the valve was installed and after the pressure was adjusted to 0.5, the patient still had poor drainage.The doctor suspected that the catheter was blocked so the abdominal catheter was removed during the surgery.It was noted that cerebrospinal fluid did not flow out of the abdominal end when the patient was in lying position.However, when the patient's head was lifted about 30 degrees, cerebrospinal fluid flew out of the abdominal end.It was noted the pressure of the valve was not adjusted during this process.As a result, the doctor suspected that the valve provided additional resistance in the supine position resulting in cerebrospinal fluid not flowing out.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STRATA II VALVE, REGULAR
Type of Device
SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS
Manufacturer (Section D)
MEDTRONIC NEUROSURGERY
125 cremona drive
goleta CA 93117
Manufacturer (Section G)
MEDTRONIC NEUROSURGERY
125 cremona drive
goleta CA 93117
Manufacturer Contact
katcha taylor
9775 toledo way
irvine, CA 92618
7635263305
MDR Report Key9508150
MDR Text Key182251062
Report Number2021898-2019-00461
Device Sequence Number1
Product Code JXG
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K042465
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/08/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number42866
Device Catalogue Number42866
Device Lot NumberF39726
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/17/2019
Initial Date FDA Received12/21/2019
Supplement Dates Manufacturer Received12/23/2019
Supplement Dates FDA Received01/08/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-