U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1 Back to Search Results
Model Number 801763
Device Problem Battery Problem (2885)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/09/2019
Event Type  malfunction  
Manufacturer Narrative
The field service representative (fsr) replaced the batteries.The unit operated to the manufacturer's specifications.The suspect parts will be returned to the manufacturer for further evaluation.
 
Event Description
It was reported that a 'battery error: battery needs service' message displayed.No other details regarding the nature of the event were provided.
 
Manufacturer Narrative
Per data log analysis, the system was powered up on (b)(6) 2019 and all looks normal.The next power up is on (b)(6)2019, and the user is notified battery life is exceeded as reported.There were no indications of a date issue that could have triggered the message.
 
Event Description
Additional information was received that the message displayed was the "you have exceeded the two year service life of your batteries" message.
 
Manufacturer Narrative
The reported complaint was confirmed.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCED PERFUSION SYSTEM 1
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
MDR Report Key9508714
MDR Text Key199262038
Report Number1828100-2019-00697
Device Sequence Number1
Product Code DTQ
UDI-Device Identifier00886799000656
UDI-Public(01)00886799000656(11)171207
Combination Product (y/n)N
PMA/PMN Number
K172220
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 02/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number801763
Device Catalogue Number801763
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/02/2020
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/09/2019
Initial Date FDA Received12/22/2019
Supplement Dates Manufacturer Received01/02/2020
02/04/2020
Supplement Dates FDA Received01/23/2020
02/27/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-