• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NEUROSURGERY STRATA II SHUNT ASSEMBLY, SMALL, WITH BIOGLIDE; SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NEUROSURGERY STRATA II SHUNT ASSEMBLY, SMALL, WITH BIOGLIDE; SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS Back to Search Results
Model Number 27818
Device Problems Insufficient Information (3190); Appropriate Term/Code Not Available (3191)
Patient Problem Therapeutic Effects, Unexpected (2099)
Event Date 12/20/2019
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information received reported that although obstruction confirmation was performed with a contrast agent, no obstruction was observed.However, the ventricle was not reduced and symptoms were not improved.It was stated there were no environmental/external/patient factors that may have led or contributed to the issue.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the post-operative progress was not favorable.It was stated the device was explanted.The patient's status at the time of the report was alive-no injury.
 
Manufacturer Narrative
The returned valve was patent.It met the requirements for valve flux, reflux, siphon and leakage testing.The catheter from (b)(6) was connected to the valve with a suture.The valve did not meet the requirements for pressure flow and preimplantation testing due to tear on the delta chamber.The instruction for use cautions, ¿improper use of instruments in the handling or implantation of shunt products may result in the cutting, slitting or crushing of components.Such damage may lead to loss of shunt integrity¿¿ dhr review of lot # e59011 did not indicate any findings that correlate to the complaint.All valves are 100% tested at the time of manufacture.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STRATA II SHUNT ASSEMBLY, SMALL, WITH BIOGLIDE
Type of Device
SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS
Manufacturer (Section D)
MEDTRONIC NEUROSURGERY
125 cremona drive
goleta CA 93117
MDR Report Key9510227
MDR Text Key182251176
Report Number2021898-2019-00464
Device Sequence Number1
Product Code JXG
Combination Product (y/n)N
PMA/PMN Number
K042465
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup,Followup
Report Date 03/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number27818
Device Catalogue Number27818
Device Lot NumberE59011
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/25/2019
Initial Date Manufacturer Received 12/23/2019
Initial Date FDA Received12/23/2019
Supplement Dates Manufacturer Received12/26/2019
03/19/2020
Supplement Dates FDA Received01/10/2020
03/23/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-