Model Number 2420-0007 |
Device Problems
False Alarm (1013); Stretched (1601); Obstruction of Flow (2423); Material Deformation (2976)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 12/02/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Although requested, product has not been received.A follow up report will be submitted with failure investigation results should the product be received for evaluation.
|
|
Event Description
|
It was reported that tubing ballooned while priming and received an occlusion alarm.There was no patient harm.
|
|
Manufacturer Narrative
|
Additional information added; h.6.(device code).The customer¿s report of balloon in the silicone segment was confirmed per photo received by the customer.Photo shows a bulge/balloon in the silicone segment tubing near the upper fitment.Prior complaints for this same failure mode determined that the ballooning is caused by excess pressure within the silicone tubing segment.The root cause for the source of the excessive pressure is unknown.
|
|
Event Description
|
It was reported that tubing ballooned while priming and received an occlusion alarm.Picture provided by the customer shows a "bulging" in the silicone segment.It was further confirmed during follow up, that there was no patient harm as a result of this event.
|
|