Model Number R SERIES |
Device Problem
Unable to Obtain Readings (1516)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
|
Event Description
|
Complainant alleged that during a routine shift check by a clinician, the device displayed a "check pads" message.Complainant indicated that there was no patient involvement in the reported malfunction.
|
|
Manufacturer Narrative
|
The customer's report was observed and attributed to a faulty top shield on the analog board.The analog board was replaced.The device was recertified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
|
|
Search Alerts/Recalls
|